Herunterladen Inhalt Inhalt Diese Seite drucken

Siemens pointek CLS 200 Kurzanleitung Seite 7

Vorschau ausblenden Andere Handbücher für pointek CLS 200:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 43
Pointek CLS 200 analog model
Specifications
For a complete listing, please see the full manual. For Approvals information, please refer to
the device nameplate.
Process Pressure/Temperature Conditions
Process Conditions
Please see
Power
• General Purpose and Explosion Proof: 12 to 250 V AC/DC 2 VA/2W max.
Ambient temperature
• general applications
• in potentially explosive
atmospheres
Wiring
WARNING: All field wiring must have insulation suitable for at least 250 V AC.
Notes:
For more detailed wiring instructions, please refer to the full manual.
Use shielded twisted pair cable; wire gauge AWG 20 to AWG 14 (0.5 mm
Maximum working voltage between adjacent relay contacts is 250 V AC.
Relay contact terminals are for use with equipment which has no accessible live parts
and wiring which has insulation suitable for at least 250 V AC.
1.
Loosen the lid clip and remove the lid to access the connectors and electronics. (The
diagram on the next page can also be found on the underside of the lid, together with a
guide to switch function).
2.
Strip the cable jacket for approximately 70 mm (2.75") from the end of the cable, and
thread the wires through the gland.
3.
Connect the wires to the terminals (polarity is not important).
4.
Ground the instrument according to local regulations.
5.
Tighten the gland to form a good seal.
6.
After adjusting the settings, replace the lid and secure the lid clip.
7ML19985QE81
on page 3.
o
–40 to 85
C (–40 to +185
check temperature class shown on device nameplate
Pointek CLS 200 (analog) – QUICK START MANUAL
o
F)
2
to 2.0 mm
2
).
Page EN-5

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis