Herunterladen Inhalt Inhalt Diese Seite drucken

Sbi Interface Protocol Descriptions - Sartorius Miras 2 Betriebsanleitung

Iw serie
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 35
SBI Interface Protocol Description
Output format with 22 characters
The following data block format is output:
I
I
I
I
I
I
I : ID
V : +/– sign
ID Codes
S t a t
Indicates a special weighing platform status, e.g., initialization, self-test,
adjustment/calibration
N
Indicates net or gross value
N 1
Indicates net value with 1
N 2
Indicates net value with 2
S / N
Indicates equipment serial number
M o d e l Indicates equipment model
Plus/Minus Sign
+
Plus sign
Minus sign
Space (i.e., no plus or minus sign; if weight value is 0 or if output does not include a
weighed value)
Unit
No stable parameter.; no weighed value
g
Grams
k g
Kilograms
l b
Pounds
o z
Ounces
t
Tons
Special Codes
The data block may contain special information.
Special status-dependent codes:
1
2
3
4
5
6
The following status codes are output for "S1" and "S2":
: Taring
C : Internal calibration
- - : All numerals shown in stable readout
H : Overload
L : Underload
V W W W W W W W W W
W : Weight value
U : Unit
st
tare memory assigned (not main tare)
nd
tare memory assigned (not main tare)
7
8
9
10 11 12 13 14 15 16 17 18 19 20 21 22
U U U CR LF
S1 S2
CR LF
23

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis