Remove sensor from operation either temporarily or permanently
10
Remove sensor from operation either temporarily or per‐
manently
User qualification: instructed user, see
Ä Chapter 2.3 „Users' qualifica‐
tions" on page 47
Disconnect the sensor electrically.
1.
Ensure that the bypass fitting is at
2.
atmospheric pressure.
Loosen the clamping nut.
3.
Pull the sensor slowly out of the bypass
4.
fitting
Unscrew the diaphragm cap and dis‐
5.
pose of the diaphragm cap.
Rinse the probes in running water until
6.
free of residue.
Loosely screw on a new diaphragm cap
7.
to protect the electrodes.
Fit the diaphragm protective cap to pro‐
8.
tect the diaphragm cap.
Store the sensor, as described.
9.
Ä Chapter 4 „Storage and Transport of
the Sensor" on page 51 .