Specifications

General Data

AC Adapter
Sartorius AC adapter
Primary
Secondary
Further data
Balance
Power supply
Input voltage
Power consumption
Ambient Conditions
These specifications apply under the following ambient conditions:
Environment
Ambient temperature *
Operational capability
Storage and shipping
Elevation
Relative humidity **
Safety of Electrical Equipment
Electromagnetic Compatibility
Defined immunity to interference
Interference emission
Standard Equipment
1
Selectable weight units
)
Available application programs
CE model balances verified for use in legal metrology comply with the requirements of Council Directive 2009/23/EC,
EN 45501:1992 and OIML R76:2006.
* For use in legal metrology in accordance with EU requirements, refer to the information on the balance.
** For use in legal metrology in accordance with EU requirements, the legal regulations apply.
1
) Depending on the country-specific model version, not all weight units listed may be available.
6971790 plus exchangeable country-specific mains plug adapter
100 – 240 V~, ±10%, 50 – 60 Hz, 0.2 A
15 V DC, ± 5%, 530 mA (max.) / 8 W (max.): 0 to +40°C and
15 V DC, ± 5%, 330 mA (max.) / 5 W (max.): 0 to +50°C
Protection class II in accordance with EN/IEC 60950-1, up to 3000 m above sea level, IP40 in
accordance with EN/IEC 60529
Only via Sartorius AC adaptor module 6971790 (package 6971790 or 6971791)
12.0–15.0 V DC
2 W (typically)
For indoor use only
+10°C to +30°C
Guaranteed between +5°C and +45°C
–10°C to +60°C
Up to 3000 m above sea level
15% to 80% for temperatures up to 31°C, non-condensing, decreasing linearly up to 50% relative
humidity at 40°C and 20% at 50°C
As per EN 61010-1 / IEC 61010-1
Safety requirements for electrical equipment for measurement, control, and laboratory use -
Part 1: General requirements
As per EN 61326-1 / IEC 61326-1
Electrical equipment for measurement, control, and laboratory use -
EMC requirements – Part 1: General requirements
Suitable for use in industrial areas
Class B (suitable for use in residential areas and areas that are connected to a low voltage network
that also supplies residential buildings). The device can therefore be used in both areas.
Gram, kilogram, carat, pound, ounce, troy ounce, Hong Kong tael, Singapore tael, Taiwan tael,
grain, pennyweight, milligram, parts per pound, China tael, mommes, Austrian carat, tola, baht,
mesghal and Newton
Secura models: counting, weighing in percent, density determination, peak hold, unstable
condition, checkweighing, mixing, totalizing, statistics, conversion, SQmin, and Identifiers
Quintix models: counting, weighing in percent, density determination, peak hold, unstable
condition, checkweighing, mixing, totalizing, statistics, conversion
Practum models: counting, weighing in percent, density determination, peak hold, unstable
condition, checkweighing