8

Technical Data

8.1 General Data

Specification
ID code (explosion protection)
Supply voltage
Input voltage
Power consumption
Further data
Ambient conditions
The technical specifications apply under the following ambient conditions:
Environment
Ambient temperature*
Operational capability
Storage and shipping
Relative humidity**
Electromagnetic compatibility
Interference resistance
Transient emissions
Available application programs
Power supply YEPS01-USB
USB power plug
Primary
Secondary
Further data
10
Installation Instructions EVO1Y | LAB1Y
Unit
Value
II 3G Ex ic nA IIB T4 Gc as per EC Type Examination Certificate no.
FM15ATEX0008X
only via USB interface or Sartorius AC adapter YEPS01-USB
V
+4.5–5.0
DC
W
2.0 (typically)
IP40 in accordance with EN 60529/IEC 60529
For indoor use only.
+10°C to +30°C
°C
Guaranteed between +5 and +40
°C
-10 to +60
%
15 to 80 for temperatures up to 30°C non-condensing, decreasing
linearly to 50% relative humidity at 40°C
In accordance with EN 61326-1/IEC61326-1
Electrical equipment for measurement, control and laboratory use –
EMC requirements – Part 1: General requirements
Basic requirements
Class B
Suitable for use in residential areas and areas that are connected to a
low voltage network that also supplies residential buildings.
Verified scales in accordance with EU requirements comply with the
requirements of Council Directive 2014/31/EC with EN 45501:2015 and
OIML R76:2006.
* For verified scales in accordance with EU requirements, refer to the
information on the scale.
** For verified scales in accordance with EU requirements, the legal
regulations apply.
Recalculation, factor calculation, formula
Type FSP007-P01P (manufacturer's designation)
100–240 V~, ±10%, 50–60 Hz, ±5%, < 0.2 A
5.2 V
, ± 5%, 1.4 A (max.)
DC
Protection class II
IP40 in accordance with EN 60529/IEC 60529