1) First of all, plug the AC adaptor
into the socket at the back of your
balance.
t
2) Then connect the AC adaptor to
the wall outlet (main voltage).

Warm-up Time

0
y To ensure accurate results are delivered,
the balance must warm up for at least
30 minutes after initial connection to the
power supply. Only after this time will the
device have reached the required operating
temperature.
30
When a verified balance of accuracy class k for use in
legal metrology is connected to the mains power, it must
warm up for at least one hour before operation.

Leveling

h
Leveling the balance compensates for slant or unevenness
at the place of installation. The balance must be perfectly
horizontal to ensure consistent, reproducible weighing
results.
Note: The balance needs to be re-leveled and adjusted
each time its setup location is changed.
Models Secura:
When "Level" appears on the display, you must level the balance.
t
Make sure that the weighing pan is
unloaded.
t
Touch "LEVEL!"
t
Follow the instructions on the
interactive display.
Models Quintix and Practum:
t
Turn the two feet as shown in the
illustration until the air bubble is
centered within the circle of the level
indicator.
– Air bubble at 12 o'clock:
turn the two feet clockwise.
– Air bubble at 3 o'clock: turn the left
foot clockwise and the right foot
counterclockwise.
– Air bubble at 6 o'clock:
turn the two feet counterclockwise.
– Air bubble at 9 o'clock: turn the left
foot counterclockwise and the right
foot clockwise.
6
Installation and Operating Instructions Secura, Quintix, Practum
Instructions for Use in Legal Metrology
Additional Display Elements:
– Metrology line with additional information about minimum load
(Min) and verification scale interval (e)
Warning symbol to indicate calculated values, negative
gross values or values that are smaller than the minimum
sample quantity (SQmin).
Setup Menu „Settings": On balances verified for use in
legal metrology, not all functions and settings are available.
Interface Specification (Data Output)
Settings „PC – SBI" and „USB RS232 SBI" for the identification of
digits not verified for use in legal metrology: In the setting „SBI"
there is no automatic identification of display digits not verified for
use in legal metrology. Corresponding measures or settings must be
carried out on the peripheral device.

Below-balance Weighing

A port for a below-balance weighing hook is located on the bottom
of the balance.
h
Install a draft protection shield.
t
Lift cover plate out of the bottom
of the balance.
Warning: When doing this, put the
balance to one side and not on the
side of the weighing pan.
t
Attach a wire (or similar) to the
sample and hang it on the notched
hook.
h
If the below-balance weighing port is no longer being
used, close the opening with the cover plate again, so that
the weighing results are not adversely affected by drafts.
The below-balance weighing port may not be opened or
used on balances used for legal metrology.

Anti-Theft Locking Device

t
A Kensington lock can be installed
at the fastening point on the back of
the balance if required.
t
The balance can also be secured at
the installation location with a chain
and lock, for example.