Excerpt: For details, see instructions for use / data sheets for the sensors used
Measuring principle
Measured value response time t
0...90
Measured value response time t
0...50
Measuring range
Zero point deviations (DIN EN 45544)
7)
Capture range
Instrument drift
Warm-up time
Effect of sensor poisons
Hydrogen sulphide H
S, 10 ppm
2
hydrocarbons, heavy metals, and
substances containing silicone,
sulphur or polymerisation
Linearity error
Standards
(Measuring function for explosion
protection and measurement of oxygen
deficiency and surplus as well as toxic
gases, DEKRA EXAM, Essen, Germany:
3)
BVS 10 ATEX E 080 X
,
PFG 10 G 001 X)
9)
Cross sensitivities
Ex
XXS O
2
catalytic
Electrochemical
oxidation
10 seconds
17 seconds for
methane
25 seconds for
propane
6 seconds
7 seconds for
methane
40 seconds for
2)
nonane
3)
0 to 100 %LEL
0 to 25 vol. %
0 to 100 vol. % for
methane
– – –
– – –
7)
7)
– – –
– – –
5 minutes
35 seconds
1 %LEL/
– – –
8 hours
Possible poisoning
5 %LEL
0.3 vol. %
8)
EN 60079-29-1
EN 50104
EN 50271
(measurement of
oxygen deficiency
and oxygen surplus)
EN 50271
10)
exist
exist
1)
XXS H
S
XXS H
S LC
2
2
Electrochemical
Electrochemical
15 seconds
18 seconds
6 seconds
6 seconds
0 to 200 ppm
0 to 100 ppm H
S
2
4)
5)
H
S
2
2 ppm
0.4 ppm
7)
7)
+/- 2 ppm
+/- 0.4 ppm
1 % of measured
1 % of measured
value / month
value / month
5 minutes
5 minutes
– – –
– – –
2 %
2 %
of the measured
of the measured
value
value
EN 45544-1/-2
EN 45544-1/-2
EN 50271
EN 50271
11)
11)
exist
exist
XXS CO
XXS CO-LC
Electrochemical
Electrochemical
25 seconds
25 seconds
12 seconds
12 seconds
0 to 2000 ppm CO
0 to 2000 ppm CO
6)
6)
6 ppm
1 ppm
7)
7)
+/- 6 ppm
+/-1 ppm
1 % of measured
1 % of measured
value / month
value / month
5 minutes
5 minutes
– – –
– – –
3 %
2 %
of the measured
of the measured
value
value
EN 45544-1/-2
EN 45544-1/-2
EN 50271
EN 50271
12)
12)
exist
exist
29