Maintenance

The instrument should be inspected and serviced once a year by suitably
qualified persons. Comparisons:
EN 60079-29-2 – Gas detectors - Selection, installation, use and
maintenance of detectors for flammable gases and oxygen
EN 45544-4 – Electrical apparatus used for the direct detection and direct
concentration measurement of toxic gases and vapours - Part 4: Guide for
selection, installation, use and maintenance
National regulations
Recommended calibration interval for measuring channels Ex, O
NO
and CO: 6 months. Calibration intervals of other gases: see Instructions for
2
Use of the respective DrägerSensors.
See the Technical Manual for details of spare parts.
Storage
Dräger recommends storing the instrument in the charger module
(order no. 83 18 639).
Dräger recommends checking the charge of the power supply at least every
three weeks if the instrument is not stored in the charger module.
Disposal
This product is not permitted to be disposed of with household waste.
This is indicated by with the adjacent icon.
You can return this product to Dräger free of charge. For information
please contact the national marketing organisations and Dräger.
Batteries and rechargeable batteries are not permitted to be disposed
of as household waste. This is indicated by the adjacent icon. Dispose
of batteries and rechargeable batteries as specified by the applicable
regulations and dispose of at battery collection centres.

Technical data

Excerpt See the Technical Handbook for details
Ambient conditions:
During operation and
storage
, H
S, SO
,
2
2
2
Temperature range over
a short period
(ATEX & IECEx only)
Position of use
Storage time
X-am 5000
Sensors
Electrical classification
Alarm volume
Operating time:
– Alkaline battery
– NiMH-
power pack:
T4
(HBT 0000)
T4 HC
(HBT 0100)
Dimensions:
Weight:
Refresh interval for
display and signals:
1)
-20 to +50 °C with NiMH power pack types:
HBT 0000 and HBT 0100,
alkaline single cell type:
2)
Duracell Procell MN 1500
-20 to +40 °C for NiMH single cell type:
2
GP 180AAHC
and alkaline single cell type:
Panasonic LR6 Powerline
0 to +40 °C for alkali single cell types:
2)
2)
Varta 4006
, Varta 4106
,
700 to 1300 hPa
10 to 90 % (to 95 % short-term) rel. hum.
-40 to +50 °C
Maximum of 15 minutes with NiMH power pack T4
2)
:
(HBT 0000) or T4 HC (HBT 0100)Requirement:
storage of the instrument at room temperature
(+20 °C) for at least 60 minutes in advance.
any
1 year
1 year
IP 67 for instrument with sensors
Typically 90 dB (A) at 30 cm distance
Typically 12 hours under normal conditions
Typically 12 hours under normal conditions
Typically 13 hours under normal conditions
approx. 130 x 48 x 44 mm (H x W x D)
approx. 220 to 250 g
1 s
27