Measurement tolerances
Accuracy in ±(% of read-off + number of digits = dgt(s) )
Accuracy is 1 year at a temperature +23°C ±5°C and relative humidity less than 75
%. Warm-up time is approx. 1 minute.
Op. mode
Meas. range
DC voltage
200 mV
2 V
20 V
200 V
1000 V
AC voltage
200 mV
2 V
20 V
200 V
750 V
DC current
2 mA
20 mA
200 mA
20 A
AC current
2 mA
20 mA
200 mA
20 A
200 W
Resistance
2 kW
20 kW
200 kW
2 MW
20 MW
Measurement voltage with measurement circuit open:
Display stabilizing time:
24
Accuracy
Resolution
10 mV
±(0.05%+3dgts)
100 mV
— — – – – — " — — – – – —
— — – – – — " — — – – – —
1 mV
— — – – – — " — — – – – —
10 mV
±(0.1%+5dgts)
100 mV
10 mV
±(0.5%+10 dgts)
100 mV
(40 Hz to 1 kHz)
±(2.5%+10 dgts)
1 mV
(1 kHz to 10 kHz)
10 mV
±(0.8%+10 dgts)
100 mV
(40 Hz to 1 kHz)
±(0.3%+3 dgts)
100 nA
±(0.3%+3 dgts)
1mA
10 mA
±(0.5%+3 dgts)
±(0.8%+5 dgts)
1 mA
±(1.5%+10 dgts)
100 nA
1 mA
(40 Hz to 1 kHz)
±(2.5%+10 dgts)
10 mA
(1 kHz to 10 kHz)
±(1.5%+15 dgts)
1 mA
(40 Hz to 1 kHz)
0.01 W
±(0.2%+5 dgts)
0.1 W
±(0.15%+3 dgts)
1 W
— — – – – — " — — – – – —
10 W
— — – – – — " — — – – – —
100 W
— — – – – — " — — – – – —
±(0.5%+5 dgts)
1 kW
< 1.2 V
2 MW approx.5s
20 MW approx.15s
Operating mode
Meas. range
Diode test
Test current 1.0 mA max.
Measurement voltage 2.8 VDC max.
beep if resistance is smaller than 30 W,
Continuity
test:
measurement voltage 2.8 VDC max.
Transistor test range: hFE 0 - 1000
: approx. 1000 mA
Base current
V
(U
)
: 2.8 V max.
CE
CE
Frequency
20 kHz
200 kHz
2 MHz
20 MHz
Input sensitivity
Max. input voltage
Maximal input parameters, overload protection
Voltage measurement ....................: 1000 VDC or 750 VAC
Current measurement ....................: 20 A AC/DC in the A-range, for 30 s max. and a
subsequent cooling phase of at least 15 min.,
250 VDC/VACrms max., overload protection:
super fast-blow 15 A 250 V fuse (dims.: 6 x 30
mm) 200mA AC/DC in the mA-range, 250
VDC/VACrms max., Overload protection: fast-
blow 0.8 A 250 V fuse (dims.: 5 x 20 mm)
Resistance measurement ..............: 20 MW, overload protection : 250 VDC/AC
Frequency measurement ...............: 20 MHz, with input voltage 250 VDC/VACrms
max.
Logic measurement .......................: 20 VDC max., overload protection 250 VDC/AC
Important!
The measurement functions transistor test and diode test/conti-
nuity check are not protected against overload or excess volta-
ge(s).
Exceeding the max. permissible input parameters may result in
damage to the measuring device and/or is a potentially lethal
hazard to the user.
Accuracy
Resolution
±(0.1%+ 2dgts)
1 Hz
— — – – – — " — — – – – —
10 Hz
— — – – – — " — — – – – —
100 Hz
— — – – – — " — — – – – —
1 kHz
: > 50 mVrms
: 250 VDC/VACrms
25