Herunterladen Inhalt Inhalt Diese Seite drucken

Ottobock 3S80 Gebrauchsanweisung Seite 23

Vorschau ausblenden Andere Handbücher für 3S80:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 5
The residual limb position, e.g., abduction, adduction and flexion, must be
considered when positioning the socket connector. Proceeding according to
the manufacturer's alignment guidelines is recommended.
Alignment is performed in 2 steps:
Bench Alignment
Static Alignment
The prosthesis and prosthetic knee joint are individually adjusted during
dynamic trial fitting.
To select and adapt the 1E90 Sprinter prosthetic foot, a trial prosthesis is
first fabricated followed by the definitive prosthesis by replacing the test foot
adapter (TF: 4R206) with the sport foot adapter (TF: 4R204).
INFORMATION
For more information on alignment, installation and optimising the pros­
thesis, please see the 647H537 Quick Reference Guide as well as the
647G839 (4R204, 4R206 TF Foot Adapter), 647G849 (1E90 Sprinter Pros­
thetic Foot) and 647G848 Instructions for Use (2Z500, 2Z501 Outer Sole).
5.1 Bench Alignment
Optimising the Prosthetic Foot Height
The sport prosthesis is fabricated as a trial prosthesis with the test foot
adapter for initial bench alignment. This verifies whether the stiffness of the
prosthetic foot was correctly selected. Then the prosthetic foot height of the
1E90 is optimised to meet the needs of the patient. Movements in the hip
and especially the shoulders offer important information on the optimum
prosthesis length. When the prosthesis length is optimised, sagging of the
shoulder on the prosthesis side while running is avoided. Further information
and exercises are found in the 647H537 Quick Reference Guide.
Bench alignment is performed with the installed 2Z500 or 2Z501 Outer Sole
(spikes not installed!).
+ = anterior placement / – = posterior placement (to the alignment reference line)
Item see fig. 4
Required tools and materials:
743A80 50:50 Gauge, alignment apparatus (e.g., 743L200
L.A.S.A.R. Assembly or 743A200 PROS.A. Assembly)
Prosthetic knee joint, locked (note the section "Use" – see
Page 27)
Bench Alignment Process
Ottobock | 23

Quicklinks ausblenden:

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

3s80 - 1 sport

Inhaltsverzeichnis