Rotors And Accessories With Limited Service Lives; Autoclaving; Centrifuge Containers - Hettich ROTANTA 460 Bedienungsanleitung

Vorschau ausblenden Andere Handbücher für ROTANTA 460:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

EN
40.2.5

Rotors and accessories with limited service lives

The use of certain rotors, hangers and accessory parts is limited by time.
These are marked with the maximum permitted number of operating cycles or with an expiration date and the
maximum permitted number of operating cycles or just with the expiration date; e.g.:
- "einsetzbar bis Ende:  V. Quartal 2011 / usable until end of:  V. Quarter 2011" or
"einsetzbar bis Ende Monat/Jahr: 10/2011 / usable until end of month/year: 10/2011"
- "Max. Lauf Zyklen / max. cycles: 40000".
For safety reasons, rotors, hangers and accessory parts may no longer be used if either the indicated
maximum number of operating cycles or the indicated expiration date has been reached.
The number of centrifuge operations can be queried; see the chapter on "Querying the operating hours and the
number of centrifuge operations".
40.3

Autoclaving

The following accessory can be autoclaved at 121°C / 250°F (20 min):

Swing-out rotors

Angle rotors made of aluminum

Hanger made of metal

Lid with biocontainment

Stands

Reductions
Otherwise you must ask the manufacturer.
No statement can be made about the degree of sterility.
The lids of the rotors and containers must be removed prior to autoclaving.
Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics.
We recommend that the packing rings of the bio-safety system be replaced after autoclaving.
40.4

Centrifuge containers


With leakiness or after the breakage of centrifuging containers broken container parts, glass splinters and leaked
centrifugation material are to be completely removed.

The rubber inserts as well as the plastic sleeves of the rotors are to be replaced after a glass breakage.
Remaining glass splinters cause further glass breakage!

If this concerns infectious material, a disinfection process is to be executed immediately.
84/203