21.2

Reset cycle counter to "0" and enter the maximum permitted number of rotor cycles

If the rotor is at standstill, this can be set as follows:
You can scroll backwards through the menu by pressing the
The operation can be cancelled at any time by pressing the
stored.

Keep the
key held down for 8 seconds.
PROG
After 8 seconds,  Machine Menu  appears in the display.

Press the
key as often as necessary until the ->Operating Time is displayed.
PROG

Press the
key. The external operating hours are displayed, e.g. OP Time ext = 0h25m.
START

Press the
key as often as is required until the rotor cycles are displayed, e.g. Cycles = 30001 of 30000.
PROG

key. The number of rotor cycles is displayed in parentheses áñ , e.g. Cycles = 30001 of
Press the
RCF
30000.

Turn the knob
to the left to reset the number of rotor cycles to "0".
If the rotor cycles are not reset to "0", press the
to be displayed and the setting is not saved.

key. The maximum permitted number of rotor cycles is displayed in parentheses , e.g. Cycles
Press the
RCF
= 0 of 30000.

Use the knob
to set the maximum permitted number of rotor cycles specified on the hanger.

Press the
key to save the settings.
START
In confirmation, Store cycles ... is displayed briefly. Next the rotor cycles are displayed, e.g. Cycles = 0 of
30000.

Press the
x 2 to quit the "Operating Time" menu or press the
OPEN / STOP
"Machine Menu".
21.3

Activating or deactivating the cycle counter

The cycle counter can be activated/deactivated as follows when the rotor is at a standstill:
You can scroll backwards through the menu by pressing the
The operation can be cancelled at any time by pressing the
stored.

Keep the
key held down for 8 seconds.
PROG
After 8 seconds,  Machine Menu  appears in the display.

Press the
key as often as necessary until the ->Operating Time is displayed.
PROG

Press the
key. The external operating hours are displayed, e.g. OP Time ext = 0h25m.
START

Press the
key as often as necessary until (with the cycle counter activated) the rotor cycles are displayed,
PROG
e.g. Cycles = 5120 of 30000, and (with the cycle counter deactivated) until Cycles = disabled is displayed.

Deactivating the cycle counter:

Press the
key as often as is required until the maximum permitted number of rotor cycles is displayed
RCF
in parentheses  , e.g. Cycles = 5120 of 30000.

Turn the knob
to the left to reset the maximum permitted number of rotor cycles to "0".

Press the
key to save the setting.
START
In confirmation, Store cycles ... and Cycles = disabled is displayed briefly.

Activating the cycle counter:

Press the
key as often as is required until the maximum permitted number of rotor cycles is displayed
RCF
in parentheses  , e.g. Cycles = 0 of 0.
Use the knob
to set the maximum permitted number of rotor cycles specified on the hanger.


Press the
key to save the setting.
START
In confirmation, Store cycles ... is displayed briefly. Next the rotor cycles are displayed, e.g. Cycles = 0 of
30000.

Press the
x 2 to quit the "Operating Time" menu or press the
OPEN / STOP
"Machine Menu".
key.
T/°C
key. In this case, the settings are not
OPEN / STOP
key, which causes Max cycles = actual cycles
START
OPEN / STOP
key.
T/°C
key. In this case, the settings are not
OPEN / STOP
OPEN / STOP
EN
key x 3 to quit the
key x 3 to quit the
73/203