Installation and removal of the rotor

Clean the motor shaft (C) and the rotor drilling (A), and lightly grease the motor shaft
A
afterwards. Dirt particles between the motor shaft and the rotor hinder a perfect seating
B
of the rotor and cause an irregular operation.
Place the rotor vertically on the motor shaft. The driver areas (D) of the motor shaft
must be located in the groove (B) of the rotor. The alignment of the groove is labelled
on the rotor.
D
C
Tighten up the clamping nut of the rotor by turning it clockwise by hand.
Check the rotor for firm seating.
Loosening the rotor: Loosen the tensioning nut by turning anti-clockwise. Turn the
tensioning nut until the rotor is able to be lifted from the motor shaft.
10

Arming the rotor

Always fill the centrifuge containers outside the centrifuge.
Ensure, by eye, that the containers are filled to a uniform level.
Load positions located opposite one another equally.
Ensure that the rotor is correctly positioned and secured.
The weight of the permissible filling quantity is specified on each rotor. This weight may not be exceeded.
11

Control panel

The control panel is divided into four areas.
Speed
Rotation display
Time
Keys
RPM x 100

Speed area

Rotation area

11.1
Speed area
The speed area contains the speed display which indicates the preselected speed or the ACTUAL speed during
operation.
The speed can be preselected or changed during operation with the
display flashes in seconds until the lid is opened or a key is activated.
11.2
Rotation area
The rotation area contains the rotation display, which illuminates and rotates anti-clockwise whilst the rotor is turning.
When the
symbols lights up, the rotor is stationary and the lid can be opened.
When the lid is unlocked, the
11.3

Time area

This area contains the running time display and the
preselected or changed during operation.
When stationary, the preselected switch-on time is displayed, as well as the remaining run time during operation. The
decimal point flashes each second during operation.
At the end of operation, the display flashes each second until the lid is opened or a key is activated.
11.4

Key area

Start key
To start a run with the preselected parameters.
START
.
All manuals and user guides at all-guides.com
START
t
STOP
Time area
Key pad
symbol lights up.
and
and
arrow keys. At the end of operation, the
arrows, with which the switch-on time can be
EN