6 Commissioning, operation
6.4.4 Setting auto shut-off parameters
The calibrator will switch itself off after a specified time; this function can also be deactivated.
Use the "AUTO OFF" option to set the auto shut-off parameters.
GB
Use the F2 and F3 keys to select the time after which the calibrator will turn itself off, or disable the
auto shut-off by scrolling down to "0".
Use the "AUTO OFF DONE" option to set the parameters and return home. The auto shut-off time
is reset whenever a key is pressed.
6.4.5 Activating and deactivating a display
Use the "DISPLAY" option on the 'Display selection' main menu to access the display activation
menu.
The F2 button can be used to select the desired display. The "ON/OFF" option switches the active
display on or off. The active display and the current status "ON/OFF" are shown in the lower
display .
Select the "DONE" option to save the changes and return home. When a display is deactivated its
configuration is retained. As soon as the display is activated, the configuration is checked against
the configurations of the other currently-active displays. If there is any conflict between
configurations, the recalled display's configuration is modified to avoid the conflict. If all three
displays are deactivated the lower display will switch on automatically.
6.4.6 Damping
Damping can be turned ON or OFF using the damping menu selection. As soon as the damping
function is switched ON, the calibrator displays the running average of the last ten measurements.
The calibrator makes approximately three readings per second.
6.4.7 Pump limit
To prevent overpressuring of sensitive devices, a maximum pressure (pump limit) can be set. In
this mode, use the arrow keys to set the maximum pressure.
WIKA Operating Instruction, Model CPH7600