Herunterladen Inhalt Inhalt Diese Seite drucken

Main Menu Functionality - WIKA Wally Box III CPH7600 Betriebsanleitung

Vorschau ausblenden Andere Handbücher für Wally Box III CPH7600:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

6 Commissioning, operation

6.1.2 Main menu Functionality

There are three options on the menu, "CONFIG", {active display} and "MORE". The main menu is
the 'home' for the menu display.
GB
6.1.2.1 Setting the {active display}
The {active display} is displayed via the middle option in the menu in the main display. Pressing the
F2 button switches between the {active display}.
6.1.2.2 Setting the {active display} parameters
In order to set the {active display} parameters, select the "CONFIG" option and open the
configuration menu.
Here the "SELECT" option will switch between the available choices for each parameter. The first
parameter is "MODE". Since voltage, current and switch test modes all use the same leads, two of
these functions cannot be used concurrently. The ability to select certain functions is limited and
dependent upon what is already selected in the other display sector. The "NEXT" option is used to
change to the second parameter. Only the pressure modes have a second parameter.
Within any single display sector, the following modes are available:
P[1] = integrated pressure sensor
[EXT] = pressure with external pressure module
P[1] ST = switch test with integrated sensor
[EXT] ST = switch test with external pressure module
Current functions are only available on the lower display:
mA measurement = current measurement without power supply
mA w/24V = current measurement with power supply
mA Source = current source
mA sim = mA simulation using an external supply from the instrument under test
VOLT = voltage measurement
The following table shows which functions are available concurrently. An "--" in a column indicates
that the mode in the active display will not be available for selection if the mode in that row is in use
in any other display sector.
P[1]
[EXT]
P[1] ST
[EXT] ST
mA
mA-loop
Volt
20
Active display
P[1]
[EXT]
P[1]
[EXT]
ST
ST
--
--
--
--
--
--
--
--
--
--
WIKA Operating Instruction, Model CPH7600
mA
mA-loop
Volt
--
--
--
--
--
--
--
--
--
--
--
--

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis