2.5 Zero correction sensor 1 (OFS.1) and sensor 2 (OFS.2)
A zero displacement can be carried out for the measured value:
Standard setting: 'off' = 0.0°, i.e. no zero displacement will be carried
out. Together with the scale correction (see below) this factor is mainly
used to compensate for sensor deviations. Input is in the display unit.
2.6 Span correction sensor 1 (SCL.1) and sensor 2 (SCL.2)
The scale of the measuring can be influenced by this setting (factor is
in %):
Standard setting: 'off' =0.000, i.e. value is not corrected. Together with
the zero displacement (see above) this factor is mainly used to com-
pensate for sensor deviations.

2.7 (P.oFF) Power off function

If no key is pressed, and no serial communication occurs during the
interval set in configuration (P.off), the device will automatically switch
itself off to save battery power.
If P.oFF = oFF then the automatic power off is deactivated.

2.8 (Out) Output

The output can be used as interface (RS-232 or USB) or as analogue
output (0 ... 1 V).

2.8.1 (Adr.) Serial communications address

This parameter has been included in preparation for future operation
with multi-point monitoring systems.
WIKA Operating Instructions Hand-Held Pressure Indicator
value displayed = value measured - offset
displayed value = measured value * (1+Scal/100)