6 Commissioning, operation

6 Commissioning, operation
As soon as the calibrator is turned on, by pressing the ON/OFF key, it will go through a short self-
test routine. During that routine, the display shows the current firmware version, auto-shutdown
status and the range of the internal pressure sensor. The calibrator requires a maximum of 5
minutes warm-up to reach its specified accuracy. Large changes in ambient temperature may
make a longer warm-up period necessary. Pressure ranges should be zeroed each time the
calibrator is started.
6.1

Calibrator display

The calibrator's display consists of two sectors: The menu bar (located along the bottom of the
screen) is used to access the menu system. The main display (the rest of the display) consists of
up to three sub-sectors for the measurement activities. These sub-sectors will be referred to as the
UPPER, MIDDLE and LOWER displays in this document.
Figure 1 shows the location of the different display fields, while are described in the following table.
1

Pressure units

Indication of the pressure unit (selectable from 13 pressure units)
2
Display of span
Indication of the 4 ... 20 mA span (only for the functions mA and mA current loop)
3
Primary parameter
Indication of the current measured parameter

6.1.1 Main display functionality

There are three options for this menu: "MENU", {active display}, and "LIGHT".
6.1.1.1 Using the "MENU" option
The "MENU" option is entry to the rest of the menu system.
6.1.1.2 Using the {active display} option
The {active display} is displayed via the middle option in the menu in the main display. It is used to
select which display on the screen, for example, the ZERO button will be used for.
6.1.1.3 Using the "LIGHT" option
The backlight can be toggled on and off using this key.
WIKA Operating Instruction, Model CPH7600
Figure 1: Display
GB