1.2 EN/IEC 60601-1-2 Table 2
The Coltolux
or end user of the Coltolux
Immunity Test
Electromagnetic
Discharge (ESD)
IEC 61000-4-2
Electrical fast
transcient/burst
IEC 61000-4-4
Surge IEC 61000-4-5 +/- 1kV differential
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power Frequency
(50/60Hx) magnetic
field IEC 61000-4-8
Note UT is the a .c . mains voltage prior to the application of the test level .
Guidance and manufacturers Declaration - Electromagnetic Immunity
LED unit is intended for use in the electromagnetic environment specified below . The customer
®
LED unit should assure it is used only in such an environment .
®
IEC 60601 test level
+/- 6kV Contact
+/- 8kV Air
+/- 2kV for power
supply lines
+/- 1 kV for input/
output lines
mode (line-line)
+/- 2 kV common
mode (line-earth)
<5% UT (>95% dip in
UT) for 0 .5 cycle .
40% UT (60% dip in
UT) for 5 cycles .
70% UT (30% dip in
UT) for 25 cycles
<5% UT (>95% dip in
UT) for 5 seconds .
3A/m
Compliance Level
+/- 6kV Contact
+/- 8kV Air
+/- 2kV for power supply
lines
+/- 1 kV for input/output
lines
+/- 1kV differential mode
(line-line)
+/- 2 kV common mode
(line-earth)
<5% UT (>95% dip in UT)
for 0 .5 cycle .
40% UT (60% dip in UT)
for 5 cycles .
70% UT (30% dip in UT)
for 25 cycles
<5% UT (>95% dip in UT)
for 5 seconds .
3A/m
Intended Electromagnetic Envi-
ronment
Floors should be wood, concrete
or ceramic tile . If the floors are
covered with a synthetic material,
the relative humidity should be at
least 30%
Mains power quality should be
that of a typical commercial or
hospital environment .
Mains power quality should be
that of a typical commercial or
hospital environment .
Mains power quality should be
that of a typical commercial or
hospital environment . If the user
of the Coltolux
LED unit requires
®
continued operation during power
mains interruptions, it is recom-
mended that the Coltolux
LED
®
unit be powered from an unin-
terruptable supply with sufficient
capacity to run the unit for the
maximum required time of the
interruption .
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital environ-
ment .