Owner's Guide
LED is green and the handpiece indicator light remains unlit,
the unit requires service . Contact your dealer or Coltène/
Whaledent Inc .
C. If the On/Off button on the handpiece does not activate the
curing light and the green light above the button is illuminated,
unit requires service . Contact your dealer or Coltène/Whaledent
Inc .
D. If your light registers a reading below 300 mW/cm
meter, take the following steps:
• Check the light lens . The light lens should be checked
daily . Replace lens with Coltène/Whaledent Inc . Light Lens
(Cat . No . C7972) if material has cured to the lens or if it is
otherwise obstructed .
• If the intensity of the light output is still under 300 mW/cm
extend the curing time according to the material's manu-
facturer recommendation until the Coltolux
be serviced . Contact your dealer or Coltène/Whaledent Inc .
for service .
E.
If activating the unit results in an error sound and the indicator
is solid orange:
• The unit requires service . Contact your dealer or Coltène/
Whaledent Inc .
VIII. REPLACING THE LIGHT LENS
IMPORTANT: The Coltolux
and must be handled with care at all times.
Never look directly into the light emitted from
the light lens or reflected from tooth or other
surfaces. Retinal damage may occur.
WARNING
Only use Coltène/Whaledent's Light Lens
(Cat. No. C7972). Do not use other manu-
facturer's light lens with this product as this
will impair performance
CAUTION
Do not touch the LED or reflector (Fig. 4)
under the light lens as hand oil contaminates
these components and can compromise
performance.
• Remove the handpiece from the charger base before
attempting to replace the light lens . Do not activate the
handpiece .
• Remove the light lens by firmly pulling it from the reflector
(Fig. 4) .
IX. INSPECTING THE LIGHT LENS
• Inspect the light lens before each use to insure that restora-
tive material has not accidentally cured to the lens . Replace
lens if material has cured to the lens or if it is otherwise
obstructed .
Never look directly into the light emitted from
the light lens or reflected from tooth or other
surfaces. Retinal damage may occur.
WARNING
on a light
2
LED light can
®
LED Light Lens is easily scratched
®
X. INSPECTING THE CHARGER BASE POWER CORD
• Periodically inspect the charger base power cord for frayed
areas or broken insulation .
XI. CLEANING
Do not immerse the unit in liquid, damage to
the unit will occur.
Do not attempt to clean the reflective surface
or the LED under the light lens as this will
cause damage to the unit or impair its
performance.
Do not saturate the light lens with cleaning so-
lution or disinfectant. Do not allow solution to
,
2
remain on the light lens for an extended period
CAUTION
of time as this may impair performance.
Do not use petroleum-based solvents, io-
dophors, or phenolic-based products on the
handpiece or charger base. (Iodophors and
phenolics can stain the surface of the unit
and damage the plastic housing.)
• All external components may be cleaned by wiping with
a soft cloth moistened with a diluted (≤70%) solution of
isopropyl alcohol .
XII. INFECTION CONTROL
Do not autoclave the unit, damage to the
unit will occur.
CAUTION
• To comply with infection control procedures, slide the
barrier sleeve over the handpiece and then attach the
light shield . A new barrier sleeve should be used for each
patient .
• The light shield should be cleaned and sterilized after each
use . Prior to use, inspect for damage . If damage is found,
discard and replace .
XIII. SPECIFICATIONS
A.
Electrical Safety
Coltolux
LED is a tested and listed device complying with
®
relevant performance, safety, and electromagnetic compatibility
requirements .
B. Electromagnetic Compatibility
See section XVI .
C. Dimensions (approximate)
Handpiece
Length: 22 .1 cm / 8 .7"
Probe Dia: 1 .22 cm / 0 .48"
Grip Dia: 2 .5 cm / 1 .00"
Charger Base
Height: 7 .4 cm / 2 .9"
Width: 6 .4 cm / 2 .5"
Depth: 8 .9 cm / 3 .5"