Ettan IPGphor 3 Safety Handbook 11-0034-61 Edition AB
1.3
Warnings
1.3.1
Instrument operation
WARNING! Ettan IPGphor 3 is a high voltage instrument that can cause fatal
electrical shock if the safety features are disabled. The safety lid must be
securely closed before starting a protocol.
WARNING! The instrument covers must not be opened by the user. It contains
electrical circuits which can give a lethal electric shock. Service and planned
maintenance should be performed by personnel authorized by GE
Healthcare.
WARNING!
Plug the instrument into a properly grounded outlet.
WARNING! The vents at the front, back, and bottom of the instrument must
not be obstructed. Air must be free to circulate for the cooling apparatus to
function correctly. There should be a minimum of 10 cm unobstructed space
behind the rear panel of the instrument.
WARNING!
Always disconnect the power cord before servicing.
WARNING! This is a Class A product. In a domestic environment, it might
cause radio interference, in which case the user might be required to take
appropriate measures.
WARNING!
Avoid spilling liquids on the body of the instrument.
volumes of liquid have penetrated the casing of the instrument and come into
contact with the electrical components, immediately switch off the
instrument and contact an authorized service technician.
Safety information
If large
1
13