S An even loading on the bearing points is to be sought. To this end, in partic
ular for statically indefinite bearings, ensure that the height level setting at
the bearing points is correct (provide spacer plates, 5 x 1 mm included in
scope of delivery).
S Undo transport protection device and stops, fit height adjustment plates
(dependent on maximum capacity, see Dimensions).
Notice
How to use the enclosed grease tube
Before fitting the load cell, the enclosed grease must be applied to the upper
thrust piece and to the underside of the pendulum bearing (with RTN...B) in
order to minimize friction between the inside surface of the thrust piece and the
load introduction element of the RTN load cell or pendulum bearing.
Use of the spacer plates provided
The scope of delivery of the module includes 2 x 5 spacer plates. They are
used for the vertical adjustment of the module. Five plates max. are to be fitted
underneath the upper module plate, and five plates max. between the base
and the cross bracket of the anti‐liftoff device
Important
Ensure that all screws are fitted with U‐washers.
RTN/M2...
A01867_02_YS0_02 HBM: public
Mounting preparation
9