OCM642 Programmierbare Widerstandsdekade
:SYSTem:PRESet
Syntax:
:SYSTem:PRESet
Description:
This command will preset all device settings. These settings are the same as the
RESET ones.
Parameters:
None
Remarks:
Overlapped command
Example:
SYST:PRES
:SYSTem:REMote
Syntax:
:SYSTem:REMote
Description:
This command places device in the REMOTE mode and locks all keys but
LOCAL key. The Command is valid only for RS232, LAN and USB interfaces.
The device will not respond to any other command until is in REMOTE mode.
Parameters:
None
Remarks:
Overlapped command
Example:
SYST:REM
:SYSTem:RWLock
Syntax:
:SYSTem:RWLock
Description:
This command places device in the REMOTE mode and locks all keys including
LOCAL key. The Command is valid only for RS232, LAN, USB interfaces. The
device will not respond to any other command until is in REMOTE mode.
Parameters:
None
Remarks:
Overlapped command
Example:
SYST:RWL
:SYSTem:TIME
Syntax:
:SYSTem:TIME <DNPD>,<DNPD>,<DNPD>
:SYSTem:TIME?
Description:
This commands allows set system device time (RTC).
Parameters:
<DNPD>
<DNPD>
<DNPD>
Remarks:
Overlapped command
Example:
SYST:TIME 10,45,15
SYST:TIME? Response: 10,45,15
57
Hours, Range 0 ... 23
Minutes, Range 0 ... 59
Seconds, Range 0 ... 59
Orbit Controls AG