OCM642 Programmierbare Widerstandsdekade
:SYSTem:COMMunicate:LAN:GATE
Syntax:
:SYSTem:COMMunicate:LAN:GATE <CPD>
:SYSTem:COMMunicate:LAN:GATE?
Description:
This command allows to change default gateway if DHCP is switched off.
Interface must be restarted to take effect (see SYST:COMM:REST command).
Parameters:
<CPD>
255.255.255.255)
Remarks:
Overlapped command
Value is not affected by reset
Example:
SYST:COMM:LAN:GATE 255.255.255.255
SYST:COMM:LAN:GATE? Response: 255.255.255.255
:SYSTem:COMMunicate:LAN:PORT
Syntax:
:SYSTem:COMMunicate:LAN:PORT <DNPD>
:SYSTem:COMMunicate:LAN:PORT?
Description:
This command allows to change port number. Interface must be restarted to take
effect (see SYST:COMM:REST command).
Parameters:
<DNPD>
Remarks:
Overlapped command
Value is not affected by reset
Example:
SYST:COMM:LAN:PORT 23
SYST:COMM:LAN:PORT? Response: 23
:SYSTem:COMMunicate:LAN:HOST
Syntax:
:SYSTem:COMMunicate:LAN:HOST <CPD>
:SYSTem:COMMunicate:LAN:HOST?
Description:
This command allows to change host name if DHCP is switched on. Interface
must be restarted to take effect (see SYST:COMM:REST command).
Parameters:
<CPD>
Remarks:
Overlapped command
Value is not affected by reset
Example:
SYST:COMM:LAN:HOST OCM642_SNXXXXXX
SYST:COMM:LAN:HOST? Response: OCM642_SNXXXXXX
53
Range
000.000.000.000
Range 0 ... 9999 (default 23)
Upper alpha, lower alpha, digits, underscores and spaces are
allowable. Maximum string size is 14 characters.
Orbit Controls AG
...
255.255.255.255
Betriebsanleitung
(default

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis