Herunterladen Inhalt Inhalt Diese Seite drucken

orbit controls OCM642 Betriebsanleitung Seite 47

Präzisions - widerstandsdekade
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

OCM642 Programmierbare Widerstandsdekade
[:SOURce]:UFUNction:CURVe:PRESet<IND_PRESET>:RCOunt?
Syntax:
[:SOURce]:UFUNction:CURVe:PRESet<IND_PRESET>:RCOunt?
Description:
This commands returns actual number of records in curve table.
Parameters:
<IND_PRESET> Range 1 ... Curve preset count (1 - if omitted)
<DNPD>
Remarks:
Overlapped command
Example:
UFUN:CURV:PRES1:RCO? Response: 2
[:SOURce]:UFUNction:CURVe:PRESet<IND_PRESET>:ROW<IND_ROW>:AMPLitude
Syntax:
[:SOURce]:UFUNction:CURVe:PRESet<IND_PRESET>:ROW<IND_ROW>:AM
PLitude <SPD>
[:SOURce]:UFUNction:CURVe:PRESet<IND_PRESET>:ROW<IND_ROW>:AM
PLitude?
Description:
This command sets / retreives selected row in curve table.
Parameters:
<IND_PRESET> Range 1 ... 255 (1 - if omitted)
<IND_ROW>
<SPD>
Remarks:
Overlapped command
Example:
UFUN:CURV:PRES1:ROW1:AMPL "10.6,220.0"
UFUN:CURV:PRES1:ROW1:AMPL? Response: "1.060000E+01,2.200000E+2"
[:SOURce]:UFUNction:CURVe:PRESet<IND_PRESET>:ROW<IND_ROW>:RDELete
Syntax:
[:SOURce]:UFUNction:CURVe:PRESet<IND_PRESET>:ROW<IND_ROW>:RDELete
Description:
This command deletes row from curve table.
Parameters:
<IND_PRESET>
<IND_ROW>
Remarks:
Overlapped command
Example:
UFUN:CURV:PRES1:ROW1:RDEL
47
Integer value representing number of records for particular
curve table
Range 1 ... 255 (1 - if omitted)
Quoted string representing amplitude. The amplitude consists
of two float numeric fields separated by comma. The first one
represents amplitude in "User Function" units and the second
one amplitude in Ohms.
Range 1 ... Curve preset count (1 - if omitted)
Range 1 ... Row count (1 - if omitted)
Orbit Controls AG
Betriebsanleitung

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis