Herunterladen Inhalt Inhalt Diese Seite drucken

orbit controls OCM642 Betriebsanleitung Seite 40

Präzisions - widerstandsdekade
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Orbit Controls AG
[:SOURce]:NICKel[:AMPLitude]
Syntax:
[:SOURce]:NICKel[:AMPLitude] <DNPD>[<UNIT>]
[:SOURce]:NICKel[:AMPLitude]?
Description:
This command sets temperature in Nickel mode. Node SOUR:NICK also selects
"NICKEL" function if not already selected. If unit parameter is part of
temperature, new unit is set.
Parameters:
<DNPD>
<UNIT>
·CEL
·FAR
·K
Remarks:
Overlapped command
Value is set to default after reset
Example:
NICK 100.0
NICK? Response: 1.000000E+02 CEL
[:SOURce]:NICKel:ZRESistance
Syntax:
[:SOURce]:NICKel:ZRESistance <DNPD>[<UNIT>]
[:SOURce]:NICKel:ZRESistance?
Description:
This command sets resistance at 0 °C for Nickel function.
Parameters:
<DNPD>
<UNIT>
·OHM
Remarks:
Overlapped command
Value is not affected by reset
Example:
NICK:ZRES 100.0
NICK:ZRES? Response: 1.000000E+02 OHM
[:SOURce]:PLATinum[:AMPLitude]
Syntax:
[:SOURce]:PLATinum[:AMPLitude] <DNPD>[<UNIT>]
[:SOURce]:PLATinum[:AMPLitude]?
Description:
This command sets temperature in Platinum mode. Node SOUR:PLAT also
selects "PLATINUM" function if not already selected. If unit parameter is part of
temperature, new unit is set.
Parameters:
<DNPD>
<UNIT>
·CEL
·FAR
·K
Betriebsanleitung
OCM642 Programmierbare Widerstandsdekade
temperature at Nickel function. Default value is 100.0 °C.
{CEL|FAR|K}
degrees of Celsius
degrees of Fahrenheit
Kelvin
Range 10.0 ... 20000.0 (default 100.0).
{OHM}
temperature at Platinum function. Default value is 100.0 °C.
{CEL|FAR|K}
degrees of Celsius
degrees of Fahrenheit
Kelvin
40

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis