Remarks:
Overlapped command
Value is set to default after reset
Example:
PLAT 100.0
PLAT? Response: 1.000000E+02 CEL
[:SOURce]:PLATinum:COEFficient
Syntax:
[:SOURce]:PLATinum:COEFficient <DNPD>,<DNPD>,<DNPD>
[:SOURce]:PLATinum:COEFficient?
Description:
This command allows to define Coefficients (A, B, C) used for "User" Platinum
standard scale.
Parameters:
<DNPD>
<DNPD>
<DNPD>
Remarks:
Overlapped command
Value is not affected by reset
Example:
PLAT:COEF 3.9083e-3,-5.775e-7,-4.18301e-12
PLAT:COEF? Response: 3.908300E-03,-5.775000E-07,-4.183010E-12
[:SOURce]:PLATinum:STANdard
Syntax:
[:SOURce]:PLATinum:STANdard <CPD>
[:SOURce]:PLATinum:STANdard?
Description:
This command selects Platinum temperature standard.
Parameters:
<CPD>
·PT385A
·PT385B
·PT3916
·PT3926
·USER
Remarks:
Overlapped command
Value is not affected by reset
Example:
PLAT:STAN PT385A
PLAT:STAN? Response: PT385A
[:SOURce]:PLATinum:ZRESistance
Syntax:
[:SOURce]:PLATinum:ZRESistance <DNPD>[<UNIT>]
[:SOURce]:PLATinum:ZRESistance?
Description:
This command sets resistance at 0 °C for Platinum function.
Parameters:
<DNPD>
<UNIT>
·OHM
41
Range 3.0e-3 ... 5.0e-3 (default 3.9083e-3), Coefficient A
Range -7.0e-7 ... -5.0e-7 (default -5.775e-7), Coefficient B
Range -5.0e-12 ... -3.0e-12 (default -4.18301e-12), Coefficient
C
{PT385A|PT385B|PT3916|PT3926|USER} (default PT385A)
Pt385 (68) standard
Pt385 (90) standard
Pt3916 standard
Pt3926 standard
User (see PLAT:COEF command)
Range 10.0 ... 20000.0 (default 100.0).
{OHM}