SCPI Commands

This chapter describes all public SCPI commands in detailed form. The commands here are
in alphabetic order.
:CALibration:RESistance:AMPLitude
Syntax:
:CALibration:RESistance:AMPLitude <DNPD>
:CALibration:RESistance:AMPLitude?
Description:
This command sets calibration value of particular internal resistance standard at
output terminals including all parasitic resistances inside the decade.
Parameters:
<DNPD>
Remarks:
This command requires "Calibration" access
Overlapped command
Value is not affected by reset
Example:
CAL:RES:AMPL 1.944
CAL:RES:AMPL? Response: 1.944000E+00
:CALibration:RESistance:SELect
Syntax:
:CALibration:RESistance:SELect <DNPD>
:CALibration:RESistance:SELect?
Description:
This command enters calibration mode and selects internal resistance standard
for calibration. Output terminals are automatically switched-on.
Parameters:
<DNPD>
Remarks:
This command requires "Calibration" access
Overlapped command
Example:
CAL:RES:SEL 0
CAL:RES:SEL? Response: 0
:CALibration:SECure:PASSword
Syntax:
:CALibration:SECure:PASSword <DNPD>
Description:
This command validates entered password and enables calibration access if
verification is successful. Acces is invalidated after reset or if CAL:SEC:EXIT
command is issued. Calibration password can be changed from the decade
system menu SETUP->Calibration->Change password.
Parameters:
<DNPD>
Remarks:
Overlapped command
Example: CAL:SEC:PASS 0
Betriebsanleitung
OCM642 Programmierbare Widerstandsdekade
Standard resistance value in Ohms. Ranges and default values
varies in accordance to selected resistance etalon (see table
"Calibration points OCM642").
Range 0 ... Max. Resistance Count -1, zero based index of
resistance standard
Range 0 ... 4294967295 (default 0)