• After scanning the upper jaw, you can continue scanning the lower jaw. On the NeoPro interface, select the
Lower Jaw
acquisition mode
OR
Press the mode button on the scanner to select the lower jaw scan mode
• Repeat steps 3 to 6 until the lower jaw scan is completed.
• Check whether there are obvious holes on the 3D model and rescan if necessary.
• When you confirm that the 3D model of the upper and lower jaw is complete, continue with the buccal bite
registration acquisition.
Scanning the buccal bite registration
To acquire a buccal bite registration, follow these steps:
• On the NeoPro interface, select the Buccal Bite Registration
OR
Press the mode button on the scanner to select the buccal bite registration mode
• Place the tip of the scanner into the buccal side in the patient's mouth, then rotate the tip to align with the
teeth, close the patient's mouth and confirm that the bite position is correct.
• Press the Start Scan button, slowly move the scanner tip in mesial direction with equal coverage of upper and
lower teeth.
• The example below shows a buccal bite registration. You can use the toolbar on the left side to switch between
the scanned occlusion views, or delete the scanned occlusion views for rescanning.
You can scan one or two buccal bite registrations. It is recommended to
scan one on the left side and one on the right side of the patient's mouth.
• After scanning the buccal bite registrations, rotate the model and zoom the view to ensure that the bite is
accurate and that there are no areas where the bite is mismatched. If necessary, you can delete the scanned
occlusion and rescan.
.
acquisition mode
.
NeoScan 1000 Installation and Operation Manual