To prepare the scanner, follow these steps:
• Make sure the lens window at the base of the scanner is clean by wiping it with a moist, lint-free cloth or lens
tissue.
• Slide the tip onto the scanner as shown below.
Starting scanning
To start scanning, place the tip of the scanner on the surface of the tooth to stabilize the scanner and press the
Start Scan button. Wait until a 3D image appears on the 3D model display screen, and then slowly move it along
the arch at 0-5mm from the teeth.
During the scanning process, adjust the surgical light to keep the light
away from the patient's mouth to avoid interference with the scanner.
Scanning approach
The recommended scanning method is to start with a molar, since it has greater details for easier identification.
Change the scanning angle to less than 60 degrees during scanning to allow the surfaces to overlap, if the
overlap is too small, the alignment may be lost.
Scanning protocol
The recommended scanning protocol consisting of 3 sweeps: occlusal, lingual and buccal to ensure good data
coverage of all surfaces.
It is recommended to start the first sweep from the occlusal surface, you should start with the first molar. The
second sweep can scan both the lingual and buccal sides, and the third scan covers the opposite side of the