Cleaning of the cutting table
bearings

Ventilating the hydraulic system

Procedure
26
Accutom-2
Instruction Manual

8. Service

If the cutting table does not move regularly at low speeds it is necessary to
clean the bearings.
Remove the 4 internal hexagon headed screws, and dismount the
aluminium plate, the water tray, the spacer tubes, the O-rings and the
lower cover plate. Notice the sequence of mounting.
Clean the bars with a lint-free cloth without solvents.
Also remove any surrounding dirt so that the bearings do not quickly
become dirty again.
Rub the bars with a thin layer of acid free oil.
Check that the cutting table can now move at low speeds.
Mount in reverse sequence.
Align the cutting table in relation to the division plate.
Fill up the holes in the hexagon headed screws with silicone matter.
Irregular, or no movement at all of the cutting table may be due to air in the
hydraulic damping circuit. Usually, this was caused during the transportation
of the apparatus.
Take hold of the left side of the apparatus so that it forms an angle of
60E with the support.
Turn off the FEED valve.
Press the FORWARD button and move the cutting table back and
forward 5 to 10 times in all its travel.
Repeat the process, this time with the FEED valve opened completely,
until the table moves satisfactorily.
Watch the pressure gauge. The pressure must not exceed max. indication.
Remember to pull the table slowly back in its starting position.