Recirculating cooling unit

6
Accutom-2
Instruction Manual
The cutting table activates a motor stop which makes automatic stop of the
cutting possible. The movement of the cutting table can be monitored on a
dial gauge.
Feeding of cooling water takes place through 2 tubes which cool and clean
the cut-off wheel on both sides.
The cooling is best carried out when connecting the recirculating cooling
unit, which belongs to the Accutom-2. At the same time, this opens up the
possibility of using special cooling agents. However, cooling can also be
carried out through connection to a water tap.
The recirculating cooling unit is made with a pump, the motor of which is
supplied with voltage from Accutom-2. This ensures synchronous start and
stop. The 12 l vessel functions as sedimentation vessel, as the pump takes
the cooling liquid from the upper layers.
The pump is resistant to water, to the current synthetic additives as well as
to oil and paraffin oil based cutting liquids.