Daily Service

Weekly Service

Monthly service

Replacing the Cooling Water

Discotom-5
Instruction Manual

5. Maintenance

Clean the cutting chamber, especially the cutting table with
the T-slots.
Clean the protection guard window with a damp cloth. Do not
use the flush hose.
Do not clean the lamp glass with alcohol. Use a damp cloth.
Empty and clean the sieve in the Recirculation Cooling Unit.
Refill the Recirculation Cooling Unit.
Always use the correct concentration of Struers Additive in the cooling
water (percentage stated on the container of the Additive). Remember to
top up with Struers Additive each time you refill with water.
Clean the cutting chamber thoroughly.
Replace the cooling water in the Recirculation Cooling Unit at
least once a month.
Place a disposable plastic insert in the tank and fold it over the
rim.
Please Note: Too high a level of coolant in the tank might
damage the pump. To avoid this, place the disposable insert so
that the pre-punched hole is in front of the overflow aperture in
the tank.
Fill the tank with 65 l of water. The water level should be 8-10
cm below the upper rim of the tank.
Add Struers Additive according to the instructions.
IMPORTANT