LaboPress-1
Instruction Manual
The noise level of the machine is 60 dB (A) measured when the
pump is running, at a distance of 1.0 m/39.4" from the machine.
LaboPress-1 is factory mounted with an electric cable. Mount a
plug on the cable:
Brown: phase
Blue: neutral
Yellow/green: earth
Check that the mains voltage corresponds to the voltage stated on the type
plate on the machine. Do not operate the machine before the transit screw
cap has been exchanged with the enclosed air filter.
Cooling water can be supplied from the water mains or a
Recirculating Cooling Unit (available separately).
Connect the pressure hose to the water mains tap for cold water:
Mount the reduction ring with gasket on the water mains tap,
if necessary.
Insert the gasket and tighten the coupling nut completely.
Water inlet
Water outlet tube
Lead the water outlet tube (pos. 2) to the drain and be
absolutely sure the hose has a steady downward slope and is
without any obstructions.
Attach the water outlet tube to the water outlet system.
An optional cooling water recirculation unit is available for users
with special environmental or other needs that demand cooling
water be recirculated. (See Accessories section.)
When connected to LaboPress-1 it must be operated manually,
using the mains power switch to start and stop the cooling
process.
Separate directions for installing and using the cooling water
recirculation unit are provided with the unit.