All components must be installed or removed as described in this manual. It is
prohibited to loosen any screws that are not mentioned in this manual, as they
appropriate torque. Unauthorized loosening of screws can weaken the system
and result in damage or injury.
!
the system will be forcefully pulled in an unpredictable manner.
The Portegra2 System can be installed to a MAVIG ceiling mounting plate, MAVIG track system, Unistrut
rails, or directly to ceiling. Prior to installation, a structural engineer must certify that the rough ceiling is
appropriate installation hardware and its torque values.
It is strongly recommended that at least two people perform the installation!
Installation of the track system, mounting plate and ceiling column, as well as the appropriate installation
hardware, must be completed according to local building codes. All local work and safety regulations must
be observed. Verify that installation complies with local requirements.
!
overhead arm to drift out of position or system to fall, resulting in operator or patient
injury. Ensure mounting surface is installed level or plumb.
!
To avoid collisions the determination of the column or track position in the room (relative
to the patient examination table) is based on following considerations: Fig. 1.1 and Fig. 1.2.
Plan positioning,
avoid collisions!
Fig 1.1 Please, carefully plan the position of the
mounting assembly. Minimize the risk of
the extension/spring arm colliding with
the column, and other equipment.
Important Safety and Precautionary Comments
No mounting
over the table!
Fig 1.2 Ensure no force can be exerted in the axial
direction within the operation area of the
extension/spring arm (at a 180° positioning).
Keep distance!
Parallel to table!
180°
®