NOTE:
It is strongly recommended that at least two people perform the installation.
NOTE:
If installing the optional MAVIG isolation kit E-TS20B05,
please follow the instructions included with the kit.
5.3.2
Carefully, slide the canopy down the ceiling column (use caution to prevent scratches in the
NOTE:
with the set screws until the installation is complete.
5.3.3
Fix the safety anchor (3/Fig. 11) to the mounting plate (4/Fig. 11) using a M8 screw and a washer.
!
Secure the chain with the attached safety latch (5/Fig. 11) to prevent it from falling
into the column tube.
5.3.4
Lift the column and engage the safety latch (5/Fig. 11) into the safety anchor (3/Fig. 11).
!
After installation, ensure the safety latch is closed to ensure the column is properly
secured.
5.3.5
Once the safety latch is engaged into the safety anchor, push up the ceiling column slightly and
rotate the safety latch (5/Fig. 11). Make sure the small diameter of the safety latch is positioned
in the safety anchor according to Fig. 11.
5.3.6
Fix the ceiling column to the mounting plate using the four M10x35 screws, lock washers, and
washers (6/Fig. 11) provided, apply Loctite
and repair of electrical components.
When installing and adjusting the Portegra2 and/or associated components, ensure
to cut off the power supply to the entire system. Danger of Death!
5.3.7
Ground the ceiling column with the screw marked with a grounding symbol (7/Fig. 11).
Complete the electrical connection to the clamp according to Fig. 12, when electricity is needed.
5.3.8
After completing the electrical connection, check again.
5.3.9
Once the ceiling column is installed, carefully slide the canopy upward until it rests against the
mounting plate and secure it with the set screws (1/Fig. 11).
Please, proceed to 7.0, „Installation/removal of plastic covers (mounting posts)", page 22.
Installing a Portegra2 column to a MAVIG mounting plate
®