When  installing  and  adjusting  the  Portegra2  and/or  associated  components,  ensure    
to  cut  off  the  power  supply  to  the  entire  system.  Danger  of  Death!
NOTE:     Connect  the  electrical  cable  of  the  cable  spooler  properly  to  the  carriage/column  unit,  before    
 
sliding  the  ceiling  column  with  carriage  into  the  track  system.  For  details,  see  drawings  (page  10).
4.2.1  
Remove  the  mounted  cable  spooler  or  end  plate  (2/Fig.4)  from  the  ceiling  track  by  loosening    
 
the  screws  (1/Fig.  4)  that  secure  it  to  the  ceiling  track.
!
Ensure  the  front  side  cover  (with  cable  spooler)  will  not  fall  from  the  ceiling  track    
when  the  last  screw  is  removed.
4.2.2  
Remove  the  small  plastic  plate  (3/Fig.  5)  on  the  top  of  the  carriage  by  removing  the  screws  (4/Fig.  5).
NOTE:  
Store  the  small  plastic  plate  (3/Fig.  5)  and  the  screws  (4/Fig.  5)  in  a  safe  place,  for  later  use.
4.2.3  
Remove  the  plastic  cap  from  the  carriage  (5a/Fig.  5)  that  points  toward  the  cable  spooler  from    
 
the  carriage  using  a  screwdriver.
4.2.4  
With  the  help  of  a  second  person,  pull  the  electrical  cable  from  the  cable  spooler  and  secure  it    
 
against  unintentional  retraction.
!
If  the  electrical  cable  is  not  secured,  the  cable  will  retract  and  the  cable  drum  will    
completely  unwind  itself,  loosing  its  preset  spring  tension.  If  that  occurs,  please    
remove  the  unit  and  replace  it  with  a  new  cable  spooler.
4.2.5  
Push  the  cable  through  the  hole  (6/Fig.  5)  that  leads  to  the  connection  clamp  on  the  carriage
 
and  complete  the  electrical  connections  according  to  „Layout  Carriage"  (Fig.  5).
4.2.6  
Secure  the  cable  with  the  tension  relief  screwed  into  the  carriage  (7/Fig.  5).
4.2.7  
Inspect  the  electrical  connection  to  ensure  its  correctness.
4.2.8  
Tighten  the  relief  cap  (5b/Fig.5)  on  the  cable  coming  from  the  cable  spooler  to  the  carriage.
4.2.9  
Mount  the  small  plastic  plate  (3/Fig.  5)  again  to  the  top  of  the  carriage  using  the  screws  (4/Fig.  5).
4.2.10  
Remove  the  cardboard  that  prevents  an  unintentional  retraction  of  the  cable.
Mounting  the  track  system  to  the  ceiling
11