4.3.1
Using light pressure, loosen the 6 screws (1/Fig. 6) of the cable spooler cover with a half turn.
4.3.2
Remove the cover (3/Fig. 6).
4.3.3
Guide the power cable from the outside through the provided pull relief cap (2/Fig. 6).
4.3.4
Complete the electrical connections according to „Layout Cable Spooler" (Fig. 6).
4.3.5
Close and tighten the cable spooler cover again, and tighten the relief cap.
4.4.1
If the front side cover is still mounted on the ceiling track, you can remove it by loosening the
counter sunk screws (8/Fig. 8) that secure the front side cover (with/without cable spooler)
to the ceiling track.
4.4.2
For Twin Column models, pull down the T-handle grip (4/Fig. 7) of the column to release the
carriage brake. If desired, you can insert the attached brake release clip (5/Fig. 7).
NOTE:
turned 90° from what is seen in Fig. 7.
For 360° Column models, pull down the T-handle grip (7/Fig. 7) of the column and turn it 90°
to the left to release the carriage brake.
Both carriage versions can only be slid along the track when the brake is released.
4.4.3
Slide all wheels of the carriage into the track.
4.4.4
After in position on the track, lock the carriage brake again.
NOTE:
4.4.5
Mount the front side cover (9/Fig. 8) back to the front side of the track. Insert the counter sunk
Please, proceed to 7.0, „Installation/removal of plastic covers (mounting posts)", page 22.
Mounting the track system to the ceiling
13