All components must be installed or removed as described. The loosening of any screws
that are not mentioned in this manual is prohibited, as they are secured with an appro-
Unauthorized loosing of screws can weaken the system and result in damage or injury.
NOTE:
The Portegra2 systems are built and validated for MAVIG original devices. Only these MAVIG
devices are optimized and equipped with all safety features. MAVIG cannot accept any warranty
or liability for non-MAVIG devices attached to Portegra2 systems, unless MAVIG has validated
and approved such devices.
!
The installation process described below is the standard process. Before
installing an end device please refer to the devices installation manual or
instructions for use for any special conditions or steps needed.
NOTE:
It is strongly recommended that at least two people perform the installation.
!
Ensure the the device will be mounted to a non-electrical spring arm.
!
Set the upper stop of the spring arm to the horizontal position (Position B/Fig. 23) during
the installation process. To correct this, insert an a 5 mm allen wrench into the opening
(1/Fig. 24) and turn it fully to the right, otherwise the spring arm will uncontrollable
swing upwards.
Installation and removal of adaptable elements
33