Important  Safety  and  Precautionary  Comments
It  is  strongly  recommended  that  at  least  two  persons  perform  the  installation.
Installing  the  system:  Prior  to  installation,  MAVIG  emphasizes  that  a  structural  engineer  must  certify  that  
the  rough  ceiling  is  strong  enough  to  provide  proper  support  for  the  entire  system  and  any  attached  end  
devices.  Installation  must  be  completed  according  to  local  building  codes.  Failure  to  follow  these  procedures    
will  release  MAVIG  from  responsibility  for  damages  of  any  sort.
Determination  of  required  installation  hardware  and  torque  values  for  installation  of  the  ceiling  column  
and  ceiling  track  is  the  sole  responsibility  of  the  structural  engineer.
Ceiling  mounted  systems  must  be  installed  properly.  Failure  to  follow  the  instructions  provided  may  lead  
to  a  potentially  dangerous  and  unstable  condition  of  the  system.  
This  product  must  be  serviced  with  MAVIG  approved,  original  replacement  parts.  Failure  to  follow  this    
recommendation  will  void  the  manufacturer's  warranty  and  will  release  MAVIG  from  all  liability  and    
warranty  claims.
When  installing  and  adjusting  the  Portegra2  and/or  associated  components,  ensure    
to  cut  off  the  power  supply  to  the  entire  system.  Danger  of  Death!
removal,  or  repair  of  Portegra2  parts  that  have  electrical  components.
!
The  Portegra2  spring  arm  is  under  constant  spring  tension.  When  installing  and    
removing  devices  to  and  from  the  spring  arm,  always  raise  it  up  slowly  to  the    
upper  stop.  Failure  to  follow  this  instruction  will  void  the  manufacturer's  warranty    
and  release  the  manufacturer  from  all  liability  and  warranty  claims.  
The  system  must  only  be  operated  by  trained  personnel.  It  is  essential  to  have  personnel  understand    
that  collisions  of  the  system  with  other  equipment  can  lead  to  severe  damages,  and  failure  of  system.
The  Portegra2  system  is  only  validated  and  approved  by  MAVIG  for  the  following  CE-­devices:  radiation  
protective  shields,  monitor  display  holders,  injector  holders,  and  Dr.  Mach  OR  and  examination  lights.
Portegra2  must  not  be  combined  with  suspension  systems  from  other  manufacturers.
 
 
 
 
 