Specifications
9
Specifications
Nominal (rated) force:
Accuracy data per ISO376 (measuring circuits connected in parallel)
Accuracy class per ISO 376
Measuring range in which the requirements of
ISO 376 are met
Reproducibility (reproducibility and repeatability
errors in varying mounting positions)
Repeatability (reproducibility and repeatability
errors in one mounting position)
Deviation from the fitting curve
Zero error
Hysteresis error (20% - 100%)
Creep
Accuracy
HBM accuracy class per VDI/VDE 2638
Relative reproducibility and repeatability errors
with unchanging mounting position
Relative reversibility error (hysteresis) at
0.4 F
nom
Non-linearity
Relative zero point return
Relative creep
Temperature coefficient of sensitivity
Temperature coefficient of zero signal
Rated electrical output
Rated output (nominal)
Relative zero signal error
Input resistance of the measuring bridges
Output resistance of the measuring bridges
Insulation resistance
Operating range of the excitation voltage
Reference excitation voltage
20
A04785_01_YCI_02 HBM: public
F
MN
nom
%
b
%
b`
%
f
%
c
f
%
0
v
%
c
%
b
%
rg
v
%
0.4
d
%
lin
d
%
s,0
d
%
crF+E
TC
%/10 K
c
TC
%/10 K
0
C
mV/V
nom
d
%
s,0
R
Ω
i
R
Ω
o
R
GigaΩ
is
B
V
U,G
U
V
ref
2
0.5
20...100
0.025
0.01
0.04
0.01
0.15
0.01
0.1
0.01
0.05
0.2
0.01
< 0.2
0.15
0.1
1.3 ... 1.7
2
755 ±1%
695 ±1%
> 2
0.5....12
5
KDB