Guidance and Manufacturer's Declarati on
The AV400, CC300, and HF470 are intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the AV400, CC300, or HF470 should assure that
it is used in such an environment.
IMMUNITY Test
IEC 60601
Test Level
Electrostati c
±6 kV contact
Discharge (ESD) IEC
±8 kV air
61000-4-2
Electrical fast
± 2 kV for power
transient/burst
supply lines
EC 61000-4-4
± 1 kV input/output
lines
Surge
± 1 kV line(s) to
line(s)
IEC 61000-4-5
± 2 kV line(s) to
earth
<5 % U
Voltage dips, short
T
(>95 % dip in U
interrupti ons and
T
voltage variati ons on
0.5 cycles
power supply input
40 % U
lines
T
(60 % dip in U
) for
IEC 61000-4-11
T
5 cycles
70 % U
T
(30 % dip in U
) for
T
25 cycles
<5 % U
T
(>95 % dip in U
T
for 5 s
Power frequency
3 A/m
(50/60 Hz) magneti c
fi eld
IEC 61000-4-8
NOTE: U
is the AC mains voltage prior to applicati on of the test level.
T
Recommended separati on distances between
portable and mobile RF communicati ons equipment and the AV400
The AV400, CC300, and HF470 are intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the AV400, CC300, or HF470
can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the AV400 as recommended
below, according to the maximum output power of the communications equipment.
Rated Maximum
Separati on distance according to frequency of transmitt er
Output
150 kHz to 80 MHz
power of transmitt er
d = 1.2√ — P
W
0.01
0.12
0.1
0.38
1
1.2
10
3.8
100
12
For transmitt ers rated at a maximum output power not listed above, the recommended separati on
distance d in meters (m) can be esti mated using the equati on applicable to the frequency of the
transmitt er, where P is the maximum output power rati ng of the transmitt er in watt s (W) according to
the transmitt er manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separati on distance for the higher
frequency range applies. NOTE 2 These guidelines may not apply in all situati ons. Electromagneti c
propagati on is aff ected by absorpti on and refl ecti on from structures, objects, and people.
-AV400_RevC_compiled.indd 47
Compliance Level
Electromagneti c environment -
guidance
±6 kV contact
Floors should be wood, concrete,
or ceramic ti le. If fl oors are covered
±8 kV air
with syntheti c material, the relati ve
humidity should be at least 30%.
± 2 kV for power
Mains power quality shouldbe that
of a typical commercial or hospital
supply lines
environment
Not applicable
Mains power quality shouldbe that
± 1 kV diff erenti al
of a typical commercial or hospital
mode
environment
± 2 kV common
mode
<5 % U
Mains power quality should be
T
(>95 % dip in U
) for
)
that of a typical commerial or
T
for 0.5 cycles
hospital environment. If the user
of the AV400 requires conti nued
40 % U
operati on during power mains
T
(60 % dip in U
) for
interrupti ons, it is recomended
T
5 cycles
that the AV400 be powered from
an uninterrupti ble power suply or
70 % U
a batt ery
T
(30 % dip in U
) for
T
25 cycles
<5 % U
T
)
(>95 % dip in U
)
T
for 5 s
3 A/m
Power frequency magneti c
fi elds should be at levels
characteristi c of a typical locati on
in a typical commercial or hospital
environment.
m
80 MHz to 800 MHz
800 MHz to 2.5 GHz
d = 1.2√ — P
d = 2.3√ — P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
Production P a er translator
Caractéristiques du produit
Guidance and Manufacturer's Declarati on - Electromagneti c Emissions
The AV400, CC300, and HF470 are intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the AV400, CC300, or HF470 should assure
that it is used in such an environment.
Emissions test
Compliance
RF Emissions
Group 1
CISPR 11
RF Emissions
Class A
CISPR 11
Harmonic Emissions
Complies
IEC 6100-3-2
Voltage Fluctuati ons
Complies
IEC 6100-3-3
Guidance and Manufacturer's Declarati on – Electromagneti c Immunity
The AV400, CC300, and HF470 are intended for use in the electromagnetic environment
specifi ed below. The customer or the user of the AV400, CC300, or HF470 should assure that it
is used in such an environment.
IMMUNITY TEST
IEC 60601 TEST
COMPLIANCE
LEVEL
Conducted RF
3 Vrms
3 Vrms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
3 Vrms
IEC 61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situati ons. Electromagneti c propagati on is aff ected by
absorpti on and refl ecti on from structures, objects, and people.
Field strengths from fi xed transmitt ers, such as base stati ons for radio (cellular/cordless) telephones
and land mobile radios amateur radio, AM and FM radio broadcast and TV broadcast cannot be
predicted theoreti cally with accuracy. To assess the electromagneti c environment due to fi xed RF
transmitt ers, an electromagneti c site survey should be considered. If the measured fi eld strength in
the locati on in which the AV400 is used exceeds the applicable RF compliance level above the AV400
should be observed to verify normal operati on. If abnormal performance is observed, additi onal
measures may be necessary, such as re-orienti ng or relocati ng the AV400. Over the frequency range
150 kHz to 80 MHz, fi eld strengths should be less than 3 V/m.
Electromagneti c Environment Guidance
The AV400 uses RF energy only for its internal
functi on. Therefore, its RF emissions are very
low and not likely to cause any interference in
nearby electronics
The AV400 is suitable for use in all
establishments other than domesti c and those
directly connected to the public low-voltage
power supply network that supplies buildings
used for domesti c purposes
Electromagneti c Environment –
LEVEL
Guidance
Portable and mobile RF communicati ons
equipment should be used no closer to
any part of the AV400, including cables,
than the recommended separati on
distance calculated from the equati on
applicable to the frequency of the
transmitt er.
Recommended separati on distance
d = 1,2√ — P
d = 1.2√ — P 80 MHz to 800 MHz
d = 2.3√ — P 800 MHz to 2.5 GHz
where P is the maximum output
power rati ng of the transmitt er in
watt s (W) according to the transmitt er
manufacturer and d is the recommended
separati on distance in meters (m). Field
strengths from fi xed RF transmitt ers,
as determined by an electromagneti c
site survey, should be less than the
compliance level in each frequency
range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
47