ENGLISH: DIRECTIoNS FoR USE
the correct tension. (Fig. L)
If the guideline is falling short of the grey block "
reapplied with more tension. (Fig. M)
If the guideline is beyond the grey block "
less tension. The patient should completely loosen the band if too much tension has
been applied before reapplying. (Fig. N)
mmHg
ANKLE CIRCUMFERENCE
26cm-30cm
STEP 5 once the bottom band has been adjusted to the correct tension, the patient
should repeat this step using the second set of guidelines located on the front of the
second band from the bottom. Be sure the patient uses the same edge
of the card and that the patient sets it to the same tension as the bottom
band. once the second band has been adjusted to the correct tension,
the patient should repeat this step for both the remaining bands, working
upwards. Ensure that the patient uses the same edge of the card and
tension setting as the other bands to assure gradient compression.
STEP 6 once the garment has been adjusted to the correct compression
level, ensure that there are no gaps between any of the bands of the
legging. Instruct the patient to roll the top part of the liner over the
top edge of the legging (Fig. o).
REMINDERS FOR THE PATIENT:
• The garment should feel firm and comfortable.
• The patient may readjust the bands as necessary throughout the day so that
all vertical guidelines are correctly matched to the appropriate pressure scale.
• Each set of vertical guidelines may be at a different location on the leg
depending on the size and shape of the leg. They do not need to be aligned
down the front center of the leg or in any other specific location on the leg.
• The Comfort Compression Anklet™ may be worn day and night, but if the
patient experiences discomfort when lying down, it should be removed.
• For improved comfort, the patient may slightly loosen the bands of the
legging for nighttime wear.
the legging should never hurt. if the patient experienCes any pain, the legging
should Be iMMediately reMoved and the patient should Consult a physiCian.
removal of the juxta-cures™ legging
To remove the Juxta-CURES™ legging, instruct the patient to detach the four bands
starting at the top band and working their way down, folding each band back onto
itself. It is important that none of the spine tabs be removed. If any portion of the spine
is accidently removed, the patient can simply reconnect it back in the same position.
", the bottom band needs to be
", the band needs to be reapplied with
mmHg
mmHg
ANKLE CIRCUMFERENCE
ANKLE CIRCUMFERENCE
26cm-30cm
26cm-30cm
If the patient is concerned with accidental spine tab removal, place the white velcro®
Stays along the edge of the spine where the patient feels the spine needs reinforcing
or trace the spine tabs with a permanent marker for easy re-application. After
removing the legging, the patient may slide the anklet and liner off the foot and leg.
follow-up visits
Remind the patient to bring the entire Juxta-CURES™ system to every follow-up
appointment for proper adjustment. At every follow-up visit with the patient, the
Juxta-CURES™ legging should be checked for necessary adjustments. Remove the
legging, take the patient's ankle and calf circumference measurements, and re-set the
spine of the legging if necessary (see steps 3-5 of the "Creating the Juxta-CURES™
Legging" section).
once the legging has been properly adjusted, re-apply the legging to the patient's
limb. Be sure to also check whether or not the patient should be using a different edge
of the Built-in Pressure System™ guide card as the appropriate range on the card may
have changed. Make the appropriate adjustments, if need be.
Tip: Once the patient's ulcer has healed, Juxta-Lite™ Leggings are available for
order to prevent ulcer recurrence. For more information, contact Customer
Service at (800) CIRCAID (247-2243) or visit us online at www.circaid.com.
this pacKage contains:
(1) Juxta-CURES™ Legging
(6) velcro® Stays
(1) Pair of Comfort Leg Liners™
(1) Pair of Comfort Compression Anklets™
(1) Built-in Pressure System™ Guide Card
(1) Paper Measuring Tape
(1) Clinician Directions for Use
(1) Patient Directions for Use
If any of these items are missing, please contact CircAid® at (800)-CIRCAID (247-
2243) or on the web at www.circaid.com.
COMPONENTS OF JUxTA-CURES™ LEGGING: Nylon, Polyethylene Plastic, Polyurethane, Spandex and Silver
COMPONENTS OF COMFORT LEG LINER™: Cotton, Lycra® and Microban® Antimicrobial
COMPONENTS OF COMFORT COMPRESSION ANKLET™: Nylon, Spandex
THE JUxTA-CURES™ LEGGING CONTAINS SILVER. THESE PRODUCTS DO NOT CONTAIN LATEx.
INDICATIONS FOR USE: venous Insufficiency, venous Stasis Ulcers, Post Thrombotic Syndrome, Dependent
Edema.
CONTRAINDICATIONS: Severe Peripheral Arterial Disease, Decompensated Congestive Heart Failure, Septic
Phlebitis, Phlegmasia Cerula Dolens, Decreased or Absent Sensation in the Leg, Allergy to Compression
Materials, Moderate Peripheral Arterial Disease, Infection in the Leg.
when bands are placed in parallel they should slightly overlap, leaving no space between bands. This
garment is not designed to be cut or modified, except as outlined in the directions for use; doing other than
instructed will void the six month limited warranty.
CircAid Medical Products, Inc. assumes no liability for injury or damages resulting from contraindicated use
of our products. CircAid is a registered trademark. Juxta-CURES, Compression Ulcer Recovery System, CURES,
Comfort Leg Liner, Comfort Compression Anklet, Built-In Pressure System and BPS are trademarks of CircAid
Medical Products, Inc. Microban is a registered trademark of Microban International, Ltd. Lycra is a registered
trademark of INvISTA. velcro is a registered trademark of velcro USA, Inc. ©2012 CircAid Medical Products,
Inc. All Rights Reserved.
ENGLISH: DIRECTIoNS FoR USE
18

Werbung

loading