7.5.2 Trouble shooting handheld

Display
Description
Regeneration is ongoing.
Battery is almost empty.
Disposable mouthpiece is not valid.
Measurement procedure is incorrect,
breathing was too strong, too weak or
time-limit after regeneration was
exceeded.
Permissible number of measurement
trials reached soon, measurements still
possible.
Number of permissible measurement
trials reached. Device does not perform
any measurements after maximum
number of permissible tests is reached.
Solution
Wait for the purging to be completed. Do not remove the
mouthpiece.
Charge the handheld on the basestation cradle, see
"3.1 Installing the handheld".
Exchange the mouthpiece and take a new one from the package,
see "4.1 Measurement using the basestation".
Repeat the measurement, see "4.2 Measurement stand-alone
with the handheld".
See display for number of remaining measurement trials. Order
new Vivatmo pro handheld, see "7.2 Status information hand-
held".
Connect a new Vivatmo pro handheld and dispose the old one,
see chapter "8.1 Disposing the Vivatmo pro device and the
battery".