Measure
Failed measurement
A) The measurement failed because the exhalation through the device was excessively weak or strong:
B) The measurement failed because the start was not performed within 60 seconds after the regeneration:
Repeat the measurement, see "4.2 Measurement stand-alone with the handheld", start at point 4.
Displaying stored measured values
Your Vivatmo pro handheld automatically stores up to 1,000 measured FeNO values, identified by time, date and measuring ID.
The measured values are stored in descending order starting with the latest.
To retrieve the measured values:
32
© Bosch Healthcare Solutions GmbH
The LED flashes red, you hear an audio signal, and the screen shows
that the exhalation failed.
Repeat the measurement, see "4.2 Measurement stand-alone with the
handheld", start at point 3.
Press the Arrow left button to show older measured values.
Press the Arrow right button to scroll to the more recent values.
Press the Enter button to exit the view.