Settings
Qualification procedure
QC Tester
Language
Jeppe M. Gregersendan
Administrator
Konsta Haapakoski
Professional
Oona Pakarinen-Kattilakoski
Professional
Jeppe M. Gregersendan
Expired
N/A
QC-Measurements
2016-04-07
11 : 43
2015-10-07
13: 16
42
© Bosch Healthcare Solutions GmbH
Note: In the QC settings, the QC Tester control must be activated, see
"6.5 Administrator settings". Perform the following steps:
Qualifiying
1. In the Settings menu, select PERFORM QC.
1/3 Passed
Qualified
The QC Tester screen opens and the actual user is highlighted and can be
selected.
2. Select the QUALIFY button.
3. Perform a regular measurement.
Q U A L I F Y
After the measurement the result can be:
• Qualify: Status during first 3 measurements of QC tester qualification.
• Passed: Result is within acceptance range of the mean of the last
3 measurements.
Passed
Passed
• Failed: Result is outside acceptance range of the mean of the last
3 measurements.
When the qualification procedure is passed, the qualification status of the
QC tester is changed to Qualified.
The qualified QC tester can perform the defined NO reference measurement test.