Settings
QC test Level 0
Jeppe M. Gregersendan
P E R F O R M Q C
Exp. Date 2018-09-10
25 PPB
QC-Measurements
2016-04-07
11 : 43
2015-10-07
13: 16
2015-06-21
09: 55
Level 0 Measurement
Last measurement : 2018-03-12 – 07:02
0 PPB
The measurement was successful.
C O N T I N U E
QC-measurement QC tester
The second reference QC-measurement is performed by a qualified QC tester. For the QC tester qualification procedure, see
"6.3.1 QC Tester qualification".
44
© Bosch Healthcare Solutions GmbH
If Level 0 QC is configured perform the following steps:
1. Select your user name and select the PERFORM QC button
(only if QC tester is activated).
2. Release the handheld device from the basestation and
Passed
attach a Level 0 mouthpiece.
Passed
Passed
3. Perform the measurement, see "4.1 Measurement using the basestation".
When the result is 0 ppb, the Level 0 QC test was successful.
If the result is 5 ppb or higher the Level 0 QC test failed.
Repeat the measurement and take care to use a new Level 0 mouthpiece.
When the Level 0 QC test still fails, contact the service.