Unassigned Result
N E W M E A S U R E M E N T
A S S I G N T O P A T I E N T
1 8 PPB
NOTE
07.04.2016 11:43
QC
Failed measurement
M E A S U R E M E N T F A I L E D
Please wait until the clock
symbol on the handheld display
has disappeared. Then press
the Enter-button to star t a new
measurement.
3. Select the x to leave the Result screen. The Patient File screen opens.
The result "5 ppb -LO-" or "<5 ppb" represents a result below the lower detection
limit of the device. The result "300 ppb - HI-" or ">300 ppb" represents a result
above the upper detection limit of the device.
When performing a measurement without selecting a patient before, the result is
shown as unassigned. You can assign the measurement to a patient immediately
or later from the list of measurements:
1. Select the ASSIGN TO PATIENT button.
The patient list is opened.
2. Select the dedicated patient from the list.
3. Measurements assigned to a patient can be unassigned by the
UNASSIGN button.
In case your measurement failed, the MEASUREMENT FAILED screen opens.
1. The screen might show being busy to prepare the next measurement. Wait until
the clock icon on the handheld stops blinking. Press the ENTER button on the
handheld and the measurement procedure starts again with the VISUALIZA-
TION screen and the animation on the screen indicates that the regeneration
process is ongoing.
The display depends on the selected visualization, see
"Visualization and regeneration" in 4.1.
Instructions for use
© Bosch Healthcare Solutions GmbH
Measure
29