Maintenance and trouble shooting
7.3

Reconditioning

Vivatmo pro performs an automatic stability check with the first measurement of the day: If the Vivatmo pro device has been
stored in warm and humid environment or not been in use for a longer time, the wrench symbol may appear on the basestation
and the handheld. Then a reconditioning is required to provide consistent measurement performance. In this case the wrench
symbol appears during the regeneration process on the pacifier screen and the device automatically starts the reconditioning.
The reconditioning takes between 7 and 90 minutes and cannot be interrupted. The wrench symbol
is shown during all activities:
70
© Bosch Healthcare Solutions GmbH
1. The wrench symbol on the handheld and the basestation indicates that the
reconditioning process is on going. The remaining reconditioning time is shown
in minutes.
2. The handheld beeps and blinks 3 times blue when reconditioning is finished.
3. Dispose mouthpiece after reconditioning.
If E-6 and the blinking battery symbol appear during reconditioning, remove the mouthpiece
and place the handheld on the basestation for 2-3h for charging. Re-attach the mouthpiece
afterwards to continue reconditioning.
Change the disposable mouthpiece if the "Invalid mouthpiece" screen appears during
reconditioning.
If you push the ON/OFF button, OFF appears as long as the reconditioning is active.