Herunterladen Inhalt Inhalt Diese Seite drucken

Switching On The Balance; Leveling The Balance; Performing An Internal Adjustment; Performing A Function Check - Mettler Toledo XPR26PC Benutzerhandbuch

Vorschau ausblenden Andere Handbücher für XPR26PC:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 33
See also
General data   Page 21

4.5.2 Switching on the balance

EULA (End User License Agreement)
When the balance is switched on the first time, the EULA (End User License Agreement) appears on the screen.
1 Read the conditions.
2 Tap I accept the terms in the license agreement. and confirm with
Warming up
Before the balance gives reliable results, it must warm up. This takes at least 120 minutes after connecting the
balance. When the balance is switched on from standby, it is ready immediately.
The balance has warmed up.
Press
.
The main weighing screen appears.
When the balance is switched on, the main weighing screen appears. The display will always show the screen
of the method last used before switching it off.

4.5.3 Leveling the balance

Exact horizontal and stable positioning are essential for repeatable and accurate weighing results.
If the message Balance is out of level appears:
1 Tap
Level the balance.
The Leveling aid opens.
2 Turn both leveling feet (1) as instructed on the display
until the dot is in the center of the level indicator.
The leveling aid can also be accessed through the Balance
menu:
Navigation:

4.5.4 Performing an internal adjustment

The adjustment Strategy is set to Internal adjustment.
1 Open the Methods section, tap
- or -
from the main weighing screen, tap
Internal adjustment is being executed.
When the adjustment has been completed, an overview of the adjustment results appears.
2 Tap
Print if you want to print the results.
3 Tap
Finish adjustment.
The balance is ready.

4.5.5 Performing a function check

To perform a function check, use the calibration kit delivered with the balance.
For further information, consult the Reference Manual (RM).
14
Balance menu >
Leveling aid
, select the adjustment, and tap
More and tap Start adjustment.
OK.
Start
u
www.mt.com/XPR26PC-RM
Pipette Calibration Balances
1

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis