Avoid direct sunlight
Sufficient spacing for balances: > 15 cm all around the instrument
Take into account the environmental conditions. See "Technical Data".
4.2 Unpacking the balance
Check the package, the packaging elements and the delivered components for damages. If any components
are damaged, please contact your METTLER TOLEDO service representative.
1 Open the box and lift the package out using the lifting
strap (1).
2 Open the lifting strap (1) and remove the User Manual
(2).
3 Remove the upper part of the package and remove the
AC/DC adapters (3) and power cables, the pump (4),
and the pipetting base with the evaporation trap container
(5).
8
Avoid vibrations
Avoid strong drafts
1
1
4
3
Avoid temperature fluctu-
ations
2
5
Pipette Calibration Balances